๐Ÿซ Lung Isolation โ€” Devices, Sizing & Position

DLT (gold std)Depth = ht/10+12FOB confirm
๐Ÿซ 4 ยท Lung Isolation

Lung Isolation โ€” Devices, Sizing & Position

Methods of lung isolation
Gold standardDouble-lumen tube (DLT)
Bronchial blocker (BB)
Single-lumen tube (SLT) โ†’ endobronchial
DLT sizing (by height)FemaleMale
< 160 cm35 Fr39 Fr
> 160 cm37 Fr41 Fr
๐Ÿ“ DLT depth & confirmation
  • Depth of DLT (cm) โ‰ˆ height (cm)/10 + 12.
  • A left-sided DLT is usually preferred (longer bronchial margin of safety); confirm placement clinically and with fibreoptic bronchoscopy (FOB) โ€” and re-check after turning the patient to the lateral position (tubes migrate).
1

Positioning โ€” lateral decubitus

  • Turn to the lateral position; re-check DLT position with FOB after the turn.
  • Avoid pressure-related complications โ€” eyes, pinna, cervical spine, hyperabduction of the arm (brachial plexus), pressure areas, the down-side neurovascular bundle.
โ† Previous๐Ÿ’ช Optimisation & Plan Next โ†’๐Ÿ’จ One-Lung Ventilation